热门搜索: 在线监测硫酸浓度设备 MPR E-Scan 安装便捷 MPR E-Scan 硫酸浓度分析仪 化工行业优选 浓硫酸在线浓度计 MPR E-Scan 数据稳定 MPR E-Scan 在线折光仪 硫酸浓度专用测量 硫酸工艺浓度控制仪 MPR E-Scan 实时检测 高精度硫酸浓度在线监测 MPR E-Scan 品牌 MPR E-Scan 硫酸在线分析仪 高温工况适用 在线硫酸浓度检测仪 MPR E-Scan 耐腐蚀探头 MPR E-Scan 折光型硫酸浓度仪 无耗材免维护 抗干扰硫酸浓度计 MPR E-Scan 在线测量 工业在线硫酸浓度分析仪 MPR E-Scan 直销 MPR E-Scan 发烟硫酸浓度仪 无惧气泡杂质 硫酸浓度在线监测仪 MPR E-Scan 精准稳定 MPR E-Scan在线硫酸浓度仪 折光原理抗干扰 MPR E-Scan在线折光仪卫生级设计食品制药适用 MPR E-Scan在线折光仪 实时数据传输 4-20mA/RS485

PRODUCT CLASSIFICATION

产品分类

新闻中心/ News Center

您的位置:首页  /  新闻中心  /  纸浆厂黑液回收锅炉在线折光仪三大故障原因分析

纸浆厂黑液回收锅炉在线折光仪三大故障原因分析

更新时间:2018-08-18      浏览次数:1357

The three major causes of refractometer trouble or failure in black liquor recovery boilers are:

Loss of cooling water and its effect on the sensing head.

Lack of reliability of the prism wash.

Condensation in the sensing head.

These may not apply to all refractometers due to differences in construction and circuitry.

Cooling Water Loss

It is of vital importance that the loss of cooling water be detected. This may be done through a temperature sensing element or flow monitor which shuts down the refractometer involved.

Damage to the sensing element of a refractometer does not occur instantaneously, but it is essential that the system detect abnormal temperatures due to cooling water loss, flow blockage, etc., and that the cooling water be promptly restored.

The individual refractometer manufacturer’s instruction and maintenance manuals shall be consulted with reference to: potential damage to the sensing element; identification of a damaged element; how and when to replace a damaged element.

Prism Wash

The time interval between prism washes may vary with the black liquor composition. It is recommended that the minimum wash period be 7-10 seconds of wash every 20 minutes. Short duration washes at more frequent intervals are more effective than long washes at long intervals. Ideally, steam pressure for prism washing should be 35 psig above the black liquor pressure, plus the pressure required to open the protective check valve.

Awareness must be maintained of the effect of changes to the prism wash programming variables. Various refractometer systems have the capability to adjust: condensate drain time, steam on time, recovery time and interval between wash time. It may be possible to configure the system to have the total time that both refractometers are in their wash cycle represent a significant percentage of operating time. If one refractometer is out of service for repairs and the remaining refractometer is in prism wash, black liquor solids are not being monitored. Prism wash should be minimized to that needed to maintain the system.

If high pressure steam is used, it may abrade the prism. If only high pressure steam is available, a reducing valve shall be used.

The refractometer prism must have a clear polished optical surface, and if it becomes abraded, it must be replaced.

If the prism wash system has not operated properly and the prism becomes coated, it must be removed and properly cleaned.

Condensation in Sensing Head

Condensate may build up in the refractometer sensing head and if this occurs, the instrument operation will be erratic.

The procedure for determining this condition and for the elimination of excessive moisture in the sensing head is not the same for all refractometers. The manufacturer’s instruction and maintenance manuals shall be consulted and followed carefully.