热门搜索: 管道化溶出MPR E-Scan在线NK浓度分析仪 MPR E-Scan氧化铝EMC Doublesensors在线Rp/αK分析仪 MPR E-Scan氧化铝溶出NK在线浓度分析仪 EMC E-Scan硫酸浓度检测仪 MPR E-Scan硫酸浓度分析仪 在线浓度分析仪在氯碱工业中的应用 在线折光仪在酚醛树脂生产中的应用 盐水浓度NaCl在线分析仪测量原理 在线上浆剂浓度分析仪在纺丝装置中的应用 在线折光仪助力己内酰胺装置硫酸浓度测量 拉丝液在线浓度分析仪工作原理及应用场景 化工抗水剂在线浓度分析仪 碳纤维油浴装置中在线油剂浓度分析仪免维护 在线折光浓度分析仪在蛋氨酸项目中的应用 在线折光仪在发烟硫酸调配装置中的应用 在线折光仪在硫酸装置中的使用及维护

PRODUCT CLASSIFICATION

产品分类

新闻中心/ News Center

您的位置:首页  /  新闻中心  /  纸浆厂黑液回收锅炉在线折光仪三大故障原因分析

纸浆厂黑液回收锅炉在线折光仪三大故障原因分析

更新时间:2018-08-18      浏览次数:1124

The three major causes of refractometer trouble or failure in black liquor recovery boilers are:

Loss of cooling water and its effect on the sensing head.

Lack of reliability of the prism wash.

Condensation in the sensing head.

These may not apply to all refractometers due to differences in construction and circuitry.

Cooling Water Loss

It is of vital importance that the loss of cooling water be detected. This may be done through a temperature sensing element or flow monitor which shuts down the refractometer involved.

Damage to the sensing element of a refractometer does not occur instantaneously, but it is essential that the system detect abnormal temperatures due to cooling water loss, flow blockage, etc., and that the cooling water be promptly restored.

The individual refractometer manufacturer’s instruction and maintenance manuals shall be consulted with reference to: potential damage to the sensing element; identification of a damaged element; how and when to replace a damaged element.

Prism Wash

The time interval between prism washes may vary with the black liquor composition. It is recommended that the minimum wash period be 7-10 seconds of wash every 20 minutes. Short duration washes at more frequent intervals are more effective than long washes at long intervals. Ideally, steam pressure for prism washing should be 35 psig above the black liquor pressure, plus the pressure required to open the protective check valve.

Awareness must be maintained of the effect of changes to the prism wash programming variables. Various refractometer systems have the capability to adjust: condensate drain time, steam on time, recovery time and interval between wash time. It may be possible to configure the system to have the total time that both refractometers are in their wash cycle represent a significant percentage of operating time. If one refractometer is out of service for repairs and the remaining refractometer is in prism wash, black liquor solids are not being monitored. Prism wash should be minimized to that needed to maintain the system.

If high pressure steam is used, it may abrade the prism. If only high pressure steam is available, a reducing valve shall be used.

The refractometer prism must have a clear polished optical surface, and if it becomes abraded, it must be replaced.

If the prism wash system has not operated properly and the prism becomes coated, it must be removed and properly cleaned.

Condensation in Sensing Head

Condensate may build up in the refractometer sensing head and if this occurs, the instrument operation will be erratic.

The procedure for determining this condition and for the elimination of excessive moisture in the sensing head is not the same for all refractometers. The manufacturer’s instruction and maintenance manuals shall be consulted and followed carefully.

关注微信公众号

邮箱:electronmachine@163.com

传真:86-21-34973725

地址:上海市陆翔路111号1号楼707室

Copyright © 2025 上海京度电子科技有限公司版权所有   备案号:沪ICP备10007481号-3   技术支持:仪表网

TEL:13761079671

关注公众号